Utvärdering av instrumentet UF-5000 för automatiserad urinpartikelanalys i en metodjämförelse med manuell mikroskopering av urinsediment
Högskolan i Jönköping, Hälsohögskolan, HHJ, Avd. för naturvetenskap och biomedicin.
Högskolan i Jönköping, Hälsohögskolan, HHJ, Avd. för naturvetenskap och biomedicin.
2018 (svensk)Independent thesis Basic level (degree of Bachelor), 10 poäng / 15 hpOppgave
Abstract [sv]

Inledning: Mikroskopisk analys av urinsediment anses i dagsläget vara gold standard inom urinsedimentdiagnostik, trots att metoden ofta är otillräckligt standardiserad. En rad automatiserade instrument, såsom UF-5000, finns idag tillgängliga och kompenserar för problematiken med den gamla metoden. Analys av urinsediment är avgörande vid flertal njursjukdomar, där bland annat erytrocyter, leukocyter, epitelceller, cylindrar och kristaller är av stort diagnostiskt värde. Syfte: Syftet var att utvärdera det automatiserade urinpartikelinstrumentet UF-5000 samt att jämföra metoden med manuell ljusmikroskopering av urinsediment. Metod: Vid metodjämförelse analyserades 69 prover med instrumentet UF-5000 samt i mikroskop. Data bearbetades med Passing Bablok-regression, Bland-Altman differensanalys samt Spearmans rangkorrelationkoefficient. Resultat: Statistisk signifikant skillnad förekom vid metodjämförelsen. Utvärderingen av instrumentet visade hög mätnoggrannhet med undantag för variationskoefficienten vid låg leukocytkoncentration samt avvikelser för instrumentets linjäritet. Diskussion: Då ljusmikroskopering av urinsediment är en otillräckligt standardiserad metod kunde statistiskt signifikanta skillnader ses, vilket kan bero på den bristande referensmetoden. Fortsatta studier där automatiserad metod jämförs med en mer standardiserad manuell mikroskopi är av stor vikt. Även reducering av cut-off för diverse parametrar kan vara aktuellt. Instrumentutvärdering visade en generell mätnoggrannhet med enstaka undantag vilket kan bero på felhantering vid analys. För vidare analys av linjäritet skulle ett ytterligare instrument som komplement kunna användas.

Abstract [en]

Introduction: Microscopy is gold standard for urine sediment analysis, although the method is often insufficiently standardized. Several automated instruments, such as UF-5000, are commercially available on the market. Urine sediment analysis is crucial for diagnosis of kidney and urinary tract diseases where elements such as erythrocytes, leukocytes, epithelial cells, casts and crystals can be identified. Aim: The aim was to evaluate the performance of UF-5000 and compare the method with manual microscopic analysis of urine sediment. Method: Analysis with UF-5000 and microscopy was conducted on 69 samples. Data were processed with Passing Bablok-regression, Bland-Altman bias plot and Spearman's rank order correlation. Result: Comparison showed statistical significance between the two methods. Performance evaluation showed high accuracy apart from the coefficient of variation for low concentration of leukocytes and systematic error for linearity. Discussion: Since microscopy of urine sediment is an insufficiently standardized method, differences were acknowledged caused by discrepancies in the reference method. Further studies where UF-5000 is compared to a more standardized manual microscopy with a larger amount of pathological urine samples is of great importance. Reduction for cut-off may also be relevant. For further analysis of linearity, an additional instrument should be used as a complement.

sted, utgiver, år, opplag, sider
2018.
Emneord [en]
Casts, crystals, kidney diseases
Emneord [sv]
Cylindrar, kristaller, njursjukdomar
HSV kategori
Identifikatorer
URN: urn:nbn:se:hj:diva-40112ISRN: JU-HHJ-BLA-1-20180071OAI: oai:DiVA.org:hj-40112DiVA, id: diva2:1216491
Fag / kurs
HHJ, Biomedical Laboratory Science
Veileder
Examiner
Tilgjengelig fra: 2018-06-21 Laget: 2018-06-11 Sist oppdatert: 2018-06-21bibliografisk kontrollert

Open Access i DiVA

fulltext(466 kB)122 nedlastinger
Filinformasjon
Fil FULLTEXT01.pdfFilstørrelse 466 kBChecksum SHA-512
a4abac411e3f150e682d3ff68097fb62217dcbde205ddd184451eb305045656154c3fa9cd49a7320f3b56c0483faabb93dc31b31472c0bff15526ab5fd8a47a9
Type fulltextMimetype application/pdf

Av organisasjonen

Søk utenfor DiVA

GoogleGoogle Scholar
Totalt: 122 nedlastinger
Antall nedlastinger er summen av alle nedlastinger av alle fulltekster. Det kan for eksempel være tidligere versjoner som er ikke lenger tilgjengelige

urn-nbn

Altmetric

urn-nbn
Totalt: 271 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf